ECG
CMRR (Common Mode
Rejection Ratio) : ¡Ã60dB
Input Impedance ¡Ã10M¥Ø
ECG Channel 1 or 2
Electrodes 3 wires
: RA, LA, LL 5
wires : RA, LA, LL, RL, V1(Chest)
Leads Select 3 Leads
: I, II, III 7
Leads : I, II, III, aVR, aVL, aVF, V1
Frequency Bandwidth
0.5~40Hz(at Filter )
Gain Selection 0.5,
1, 2, 4 mV/cm
Heart Rate Pulse Range
30~300 bpm
QRS Indication Available
Heart Rate Accuracy
¡¾1bpm or 2%
Isolation Breakdown
Voltage >4000V AC 60 Sec 60Hz
Patient Leakage Current
¡Â100uA
Notch Filter 50/60
Hz
Defibrillator / ESIS
Protected
Respiration
Excitation Current
< 500uA
Excitation Frequency
50kHz
Measuring Method
Thoracic Impedance Detection between RA and LA of ECG Electrode
Measuring Range
2~99 bpm
Accuracy ¡¾1bpm or
2%
Sensitivity 0.2~4¥Ø/cm
Temperature
Measuring Range
15.0~45.0¡É
Probe Compatible
YSI probe and Exacon
Scale Celsius Degrees
Accuracy ¡¾0.1¡É
Power
Direct Current Voltage
6.5~8.0V
Consumption Current
Maximum 190mA at 7.5V
Dimension 120¥ª68¥ª16mm(W¥ªD¥ªH)
|